Home > Support > SR245 Tech Support

SR245 Tech Support

If you are experiencing trouble with your SR245 you can either send your instrument back to SRS for service, or you can try and fix the problem yourself. If you wish to send the instrument back now click here.

If you wish to try and fix the problem yourself, the following troubleshooting guide may help you.

SR245 Troubleshooting Guide

Symptom 1 — Can't send commands remotely over GPIB interface

Click here to download programming instructions using National Instruments Measurement & Automation Exporer.

Symptom 2 — Can't send commands remotely over GPIB interface

A modification to the GPIB interface of the SR245 Computer Interface Module was implemented in June 1991. A 270 pF capacitor was added between pin 10 (ground) and pin 17 (DAC) on the GPIB bus driver (part number 75161, adjacent to the MC68488 controller). If you have an old unit, check to see if there is a capacitor installed between pins 10 and 17 of the 75161 chip (located near the GPIB connector). If a 270 pF capacitor isn't soldered between these pins, you will need to install one before GPIB communication is possible.

_____________________________________________________________________________________

If you are still having trouble your SR245, it will need to come back to SRS for service. Click here to get an RMA number.