Search
Home > Products > Vacuum & Analytical Instruments > NL100

Nitrogen Laser NL100 ... from $4750
NL100 — 170 µJ N2 laser
(Click to enlarge)
NL100 Nitrogen Laser
  • 337 nm wavelength
  • 170 µJ pulse energy
  • Internal or external triggering to 20 Hz
  • Replaceable laser cartridge
  • No mirror alignment necessary
  • TTL level sync output pulse (opt.)
  • Fully compatible with VSL-337i OEM
The NL100 Nitrogen Laser is ideal for fluorescence measurements, MALDI-TOF mass spectrometers, and other pulsed UV radiation experiments. It provides 3.5 ns pulses at 337 nm (UV), with repetition rates up to 20 Hz. The pulse energy is 170 µJ, which results in a peak power of 45 kW and an average power of 3 mW.
   Related Products

> RGA Residual Gas Analyzers
> Bayard-Alpert Gauges
> PG108 Pirani Gauge
Downloads

Datasheet
User's Manual
Prices

NL100 Nitrogen Laser

The NL100 Nitrogen Laser is ideal for fluorescence measurements, MALDI-TOF mass spectrometers, and other pulsed UV radiation experiments. It provides 3.5 ns pulses at 337 nm (UV), with repetition rates up to 20 Hz. The pulse energy is 170 µJ, which results in a peak power of 45 kW and an average power of 3 mW.

The NL100 can be triggered internally or externally at rates up to 20 Hz. It can also provide a sync output pulse (optional) derived from the laser pulse for experiments or systems where sub-nanosecond accuracy is critical. The user also has the option of running the laser system in command charge mode.

The NL100 uses a replaceable, sealed laser cartridge which includes the high voltage storage capacitors, switching element, and laser tube. The cartridge is warranted to maintain at least 70 % of its energy for twenty million pulses or one year, whichever occurs first.

No mirror alignment is ever necessary in the NL100, as the laser optics are mounted on the plasma tube and aligned at the factory. The NL100 also includes all safety features necessary to comply with the U.S. laser safety standards contained in 21 CFR 1040.10.

The NL100 is fully compatible with the Spectra-Physics model VSL-337i OEM Nitrogen Laser (part #337999-04). The mechanical design of the NL100 includes all of the mounting and alignment features of the VSL-337i, making the NL100 a straightforward retrofit in many pre-existing systems.

 

 

 

 

Phone: (408)744-9040 • Fax: (408)744-9049 • email: info@thinkSRS.comwww.thinkSRS.com

NL100 Nitrogen Laser
Beam Characteristics
Wavelength 337.1 nm
Spectral bandwidth 0.1 nm
Pulse width (FWHM) <3.5 ns
Pulse energy 170 µJ
Energy stability
(pulse to pulse)
3 % std. dev. (at 10 Hz)
Peak power 45 kW
Average power 3 mW (at 20 Hz)
Beam size 3 × 7 mm
Beam divergence
(full angle)
5 × 8 mrad
Triggering
Repetition rate 0 to 20 Hz (external trigger)
1 to 20 Hz (internal trigger)
External trigger input TTL (opto-isolated)
Internal trigger generator 1 to 20 Hz (adjustable)
Sync output pulse TTL level (opt.)
General
Power requirements +24 VDC, 1.5 A (average) at 20 Hz., 3 A (peak)
Power consumption 36 W (20 Hz operation)
Operating temperature <35 °C
Key switch On/off
Interlock switch Built-in
Dimensions 3.75" × 3.75" × 11" (WHL) (9.5 cm × 9.5 cm × 27.9 cm )
Weight 7.5 lbs., 3.4 kg
Warranty Cartridge is warranted to maintain at least 70 % of its energy for twenty million pulses or one year, whichever occurs first.
 
 
 
 
Phone: (408)744-9040 • Fax: (408)744-9049 • email: info@thinkSRS.comwww.thinkSRS.com